Surgical Instrument Sterilization Guidelines
- Imprimir
- OscuroClaro
- DF
PPE
Required PPE
|
Precleaning
What | Where | How |
---|---|---|
McKesson Multi-Enzymatic Foaming Cleanser Spray | Exam room |
|
Transport
All soiled items must be transported in a puncture-resistant, locked container
Cleaning and Drying
What | How |
---|---|
Dilute solution | McKesson Multi-Enzymatic Cleanser diluted with water
per
|
Soak |
|
Rinse |
|
Dry |
|
Dispose of cleaning solution |
|
Packaging
What | How |
---|---|
Inspect | Inspect instruments for damage and place hinged instruments in the open position |
Apply | Apply instrument lubrication |
Package | Place in self-sealing pouches or autoclavable procedure trays Do not overlap instruments Lot label must be placed on outside |
Sterilization
Important: Hinged instruments must be open when placed in the package for sterilization
Brand | Instrument | Autoclave Setting | Sterilization Time (Minutes) | Sterilization Temperature | Drying Time (Minutes) |
---|---|---|---|---|---|
BR Surgical | Forceps: Tissue, Hemostatic, Kelly, Biopsy, Splinter, Tenaculum | A | 4 minutes | 270°F (132°C) | 30 minutes |
Iodine Cup | |||||
IUD Extractor Hook | |||||
Kevorkian Curette | |||||
Nail Splitter | |||||
Needle Holder | |||||
Scissors | |||||
Sterilizer Tray | |||||
Sims Uterine Sound | |||||
Argent McKesson | Tubular Cage Finger and Toe Applicator |
Storage
Sterile instruments and supplies should be stored in covered or closed cabinets.
Sterile instruments and supplies should not be stored under sinks or in other places where they might become wet.
Use (including inspection for damage)
Instruments are to be inspected before opening and used to ensure that the packaging has not been compromised (i.e., opened, wet, torn, or punctured) during storage. If the packaging has become damaged in any way, the instruments must be reprocessed.
Wrapped packages of sterilized instruments do not have a formal expiration date or shelf life.
Instruments should remain sterile until an event causes the item to become contaminated.
All labels indicate “indefinite shelf life unless damaged, opened, or wet.”